MRI Test Objects
|
MRI Test Objects - the "Eurospin II Test System"
These Test Objects are designed to be used for checking the image performance of MR Scanners with static fields in the range from 0.02 to 3 Tesla. The design was based upon Collaborative Research work within the framework of the Concerted Research project "Identification and Characterization of Biological Tissues by NMR" COMAC-BMU II 2.3 from 1984 led by Dr F Podo. It was further developed by a group led by Dr R A Lerski at Hammersmith Hospital Medical School. |
|
 |
| Loading Annulus for the Eurospin II Test System |
 |
| The Eurospin II Test System for MRI Scanners |
|
- Test Object TO1 - UNIFORMITY - is for assessing Uniformity, Ghosting, Image Signal-to-Noise Ratio and Uniformity of Image Signal-to-Noise Ratio.
- Test Object TO2 - SLICE - is for assessing Geometric Distortion, Slice Profile and Slice Width.
- Test Object TO3 - SLICE POSITION - is for assessing Slice Warp and Slice Position (single and multisplice).
- Test Object TO4 - RESOLUTION - is for assessing Resolution by Bar Pattern and the Modulation Transfer Function.
- Test Object TO5 - CONTRAST - is for measuring T1 & T2 Precision and Accuracy, and Image Contrast. TO5 is supplied with a set of 18 Gel-filled Test Tubes.
- The LOADING ANNULUS is an optional accessory device into which the various Test Objects are inserted to simulate the body loading.
|